Lab 6.10 — AGC/LNA Gain-Transition Phase-Discontinuity Analysis

Goal

The existing two-board QPSK link (Block 11) deliberately uses fixed RX gain. This lab turns on AD9361/AD9363 AGC on purpose and measures what a gain-state transition does to carrier phase, separated honestly from the ordinary phase rotation CFO already produces every sample.

A receiver's phase keeps turning continuously because of CFO even with no gain event at all. A raw "phase before vs phase after" comparison at a gain step cannot tell a gain-correlated phase glitch apart from that ordinary rotation: the two must be separated by fitting the CFO trend and looking at the residual at the transition.

This complements Lab 6.2 and issue #29 (AD9363 gain table and overload limits), which characterize static gain/overload behavior. This lab is about dynamic gain transitions.

Method

fixed-gain baseline
    -> enable AGC
    -> force/observe a gain-state transition
    -> fit the pre-transition CFO trend (phase vs sample index)
    -> extrapolate the trend across the transition
    -> discontinuity = observed phase - extrapolated phase
    -> amplitude step, recovery time, EVM/BER impact

Executable analysis tool

blocks/block_06_rf_frontend_and_ad9363/python/lab_6_10_agc_phase_transition_analysis.py

Run the no-hardware self-test first:

python blocks/block_06_rf_frontend_and_ad9363/python/lab_6_10_agc_phase_transition_analysis.py --self-test

The self-test builds two synthetic captures sharing the same injected CFO and the same amplitude step: one with an additional injected phase discontinuity at the transition, one without (a control). A PASS means the tool recovers the injected discontinuity to within 0.05 rad on the first capture and reports near-zero on the control: i.e. it does not mistake ordinary CFO rotation for a gain-transition event. It proves the analysis method, not any real AD9361/AD9363 behavior.

For a real capture:

python blocks/block_06_rf_frontend_and_ad9363/python/lab_6_10_agc_phase_transition_analysis.py \
  measurements/agc_001.ci16 --metadata measurements/agc_001.json \
  --output measurements/agc_001_analysis.json

The metadata sidecar follows templates/gain_transition_capture_metadata.template.json: sampling format/rate and a gain_transitions list of sample_index values (plus the gain-control mode and settings for the report). Every listed transition is analyzed independently.

Procedure

The phase fit requires a tone or IQ with the known symbol modulation removed. Do not fit a straight CFO trend to raw QPSK data phase. recovery_samples is the first run of 16 samples within 0.15 rad of the pre-event phase trend; null means no such run was observed within the search window. A permanent phase step does not recover. This measures signal-phase return, not carrier-loop recovery; modem recovery needs the separate demodulator experiment. File input is labelled unverified-capture-analysis until its provenance is reviewed.

  1. Baseline with fixed RX gain. Reproduce the existing two-board cabled QPSK link (or a simpler tone/loopback bench). Record RX gain, attenuation, RSSI/dBFS, CFO, EVM and BER, and save a reference IQ capture with no gain transitions.
  2. Enable AD9361/AD9363 AGC. Document the selected gain-control mode and relevant register/iio_attr settings. Apply controlled input-level changes with a digital or fixed attenuator to force transitions across gain-table regions/LNA states where practical.
  3. Capture the transition. Save raw complex IQ before, during and after each gain change, with enough samples on each side for the CFO fit (--fit-window, 256 samples by default). Record the actual sample_index of each transition in the metadata.
  4. Run the analysis tool on the real capture and record its amplitude_step_db, phase_discontinuity_deg and recovery_samples per transition: do not hand-estimate these from a plot.
  5. Evaluate modem impact. EVM/BER/PER around the transition, frame-sync loss/reacquisition if it occurs, comparison against the fixed-gain baseline.

Report

Use templates/agc_phase_transition_report.template.md. It requires, among other fields: exact gain-control settings, the attenuation/input-level sweep, a raw/replayable IQ capture reference, the required plots (gain-state/level vs time, phase error vs time with the fitted CFO trend overlaid, constellation before/during/after), and an explicit statement of whether the observed phase behavior is deterministic, state-dependent or variable across repeats.

What this lab does not claim

--self-test is a synthetic proof of the separation method (CFO trend vs transition-correlated residual), not a hardware result. Real AGC phase-discontinuity evidence requires a real gain-state transition captured on AD9361/AD9363 hardware: this environment does not have that hardware, so no such evidence is claimed here. The lab is hardware-pending until a real capture is analyzed and reported through the template above.